The
diffractogram or diffraction pattern


This diagram illustrates the (-114) plane of the calcite structure.
XRD data are sensitive to as many as 35 different factors.
These factors can be grouped in to three
general sources of "error". The word error is used here in the
context of what causes deviations from theory. Sources of error
include:
1. Instrument sensitive errors
2. Sample sensitive errors
3. Specimen sensitive errors
Theoretical d-value (nλ = 2d sinθ)
vs.
Practical d-value (theoretical + inherent aberrations)
vs.
Experimental d-value (practical + inherent sample aberrations + errors)
Geometry of
the powder diffractometer
The essential features of the powder method includes a narrow beam of monochromatic X-ray radiation impinging upon a randomly oriented powder, which has all possible crystallographic planes available for Bragg reflection.
Geometrical
Principles of a Bragg-Brentano Parafocusing Diffractometer
1. X-rays diverge from source.
2. The "reflected" X-rays from the samples on the focusing circle are directed to their respective places back on the focusing circle.
3. The spots labeled G1, G2, G3 are the respective reflections of d-spacings d1, d2, d3.
The Bragg-Bentono geometry allows for a constant distance between the sample and the detector.
The geometry requires that the distance from the source to the sample and the sample to the detector be equal (i.e., R1 = R2) and the sample is kept on the tangent of the focusing circle.
In order to keep the detector distance constant, the sample must rotate at 1/2 the angular velocity of the detector. As the angle of incidence (θ) changes, the detector must move 2θ . This is called 2:1 motion.
Instrument alignment. All alignment steps rely on the fact that the sample is always equa-distant between the source and detector. (i.e., R1-incident = R2-reflected)
a. Axial alignment: The beam path must make a
straight line from the source (F) to sample (S) to receiving
slit (G) at the detector.



c. Zero alignment: The source, the sample, and the detector must all be perfectly in line at 0° θ.

d. Alignment of slits. The divergence slit limits the total irradiation area of the sample. The aperture of this slit hardly affects relative peak intensities if the slit is fixed and the specimen completely intercepts the X-ray beam. Sample mounts are typically hold material in 2.5 cm wide by 2.5 cm long holder.

The figure above shows both the theoretical and measured the relationship between sample irradiation length versus 2θ for several different divergence slits. The raduis of modern diffractometers ranges from 180 to 250 cm. Notice that for a 1° slit that a length of 2.5 cm is reached at 13° 2θ. At lower angles the sample intercepts only part of the beam resulting in reduced beam intensity per unit area of sample surface and increases the likelihood of background scatter from the edges of the sample holder. At higher 2θ angles less area is irradiated, which would have an effect of decreasing diffraction intensity. The depth of penetration of the beam becomes commensurably deeper with higher angles. This effectively increases background as well as a sample displacement effect (see below). Theta compensating divergence slits on some diffractometers are designed to lessen high-angle intensity loss and displacement effects and low-angle background scattering. One must always recall however, that the standard form of reporting relative intensity is with fixed-slit conditions and that variable slit data must be appropriately corrected.
Five
instrumental functions
![]()

An addition nuance, regarding beam voltage
and current involves that fact that the tube filament may change
location with different settings. The position of the filament
at low voltage and current may not be the same at high voltage
and current. This can have consequences for alignment and
other procedures that depend on accurate and precise intensity
values from the tube.
2. Flat specimen error

Note that the diverging path lengths of the
above beam are both longer and shorter than the center bram path
(i.e., as it spreads out over the sample). Ideally, we would
like the sample to sit as curved surface on the focusing
circle. It is impracticle to create such a curved surface
(it's easy and more reproducible to make a flat surface), so we
accept a small amount of distortion of the signal due to flat
specimen that sits tangential to the circle.
3. Axial divergence

There is additional opportunity for the beam
to diverge in the horizontal plane. The above figure shows that
beam path can only be longer than the ideal centered beam.
Closely spaced foil films oriented parallel to the center beam
absorb the more widely diverging beam paths. These are call
soller slits. Each instrument manufactuer has a varient for how
to optimally minimize this affect. Here
is one example.
4. Penetration
of
the
beam
into
the sample. (Sample transparency). The more intense the incident
radiation, the farther the beam penetrates into the sample. This
creates a situation similar to flat specimen error as well as
penetration into the substrate holding the sample. In the latter
case, diffraction from the substrate (off the focusing circle)
will contribute to the pattern.

5. Receiving slit - Increasing the width of the receiving slit generally increases the peak height and width and decreases the ability to resolve peaks.
The net affect of instrument line profile modifications is to broaden and displace the theoretical line position to a lower theta angle. Kα2 radiation will displace observed lines to higher angles.
6. A sixth function can be considered
as general misalignment. The figure below is reproduced from
Klug and Alexander (1974). It shows the six functions and their
weighting. Each is convoluted to form the final line profile.
The sum of all sources of error results in the low angle bias to
the peak position. This is distortion can be measured
using reference standards.

Detector dead time
When the measured count rate is not directly proportional to the photon rate entering the detector, the detector response is non-linear and said to have dead time. The effect is to increase the relative intensity of the weaker peaks. Most modern detectors (with calibration and computer correction) handle count rates up to 100,000 counts per second (cps). A quick check to determine if the detector not responding in a linear fashion is to measure the intensity of the strongest line with the tube current set for the normal quantification. Perform the measurement a second time at half the tube current. If the second measurement is more than half the first then detector dead time is the probable cause. In this latter case the detector correction routing must be recalibrated or the observed data must be corrected accordingly.

The precision of intensity data can be limited more by counting statistics than any other single parameter except preferred orientation. The figure below illustrates how the probable error of peak intensity measurements can vary with both the total number of counts and the ratio of the peak to background counting rate (R).

Both the total number of counts and R depend upon scan rate and chop or step increment. For a chop increment of 0.01° and scan rate of 2° 2θ min-1, the count time per increment is only 0.3 s. If the counting rate at the peak top is 1000 cps, then the total number of counts is 300. If R is large, then the best precision attainable is about +4%. If R is 1.5, then probable error exceeds +10%. By decreasing the scan rate of 0.2° 2θ min-1, probable error drops to +5%.

Further improvement in peak intensity precision is possible by smoothing over the peak top in small increments and also averaging background over hundreds of increments on both sides of a peak.
Sample displacementDisplacement of the sample off the diffractometer focusing circle can be brought about in three ways. Firstly, there is instrumental misalignment. Often neglected is the tacit assumption that the goniometer is properly aligned before any experiment is run. This can be easily maintained with a proper alignment, using manufacturer provided tools and checking with a standard material such as the U.S. National Institute of Standards and Technology Reference Material 640b.
If the sample holder is properly aligned, then the second potential source of displacement error comes from way the sample itself is packed into the sample holder (see next section).The figure below shows the changes in measured d-spacings for a series of reflections as a function of displacement from the goniometer focal plane.

Displacement error increases rapidly as 2θ falls below 20° (see in the figure below).

Equally important is the decrease in
diffraction intensity with increasing displacement. Finally,
effective displacement or sample transparency can be a
consequence of a low mass absorption coefficient (µ) or
high sample porosity (see next section).
Background
Background can be produced from a number of sources. These include:
1. Fluorescent radiation emitted by the specimenPeak fitting procedures typically are needed to consider removal of background. Without correction for the background much of the trace is fit with nonsensical peaks that do not provide a unique solution Most XRD manufacturers provide peak fitting software to allow this procedure.

The figure above demonstrates a single peak fitting approach. The form of the fit is:
Intensity = baseline + Kα1 Gaussian
(peak ht. and width) + Kα2 Gaussian (peak ht. and
width).
Specimen sensitive errors are most commonly introduced while preparing the sample for presentation to the X-ray beam. Itemized below are the most profound parameters related to preparation and mounting of powder.
1. Coherent scattering domain.2. Preferred orientation
The XRD powder method relies on the principle that all possible crystallographic orientations are presented to the beam. This concept is known a random orientation. If there is a bias of orientations of one or more particular crystallographic plane, then this is known a preferred orientation. Preferred orientation is likely the most common cause of intensity variations in XRD powder experiments.
Particles with perfect cleavage or acicular shapes, such as clay minerals, are the most prone to preferred orientation. Intensity variations up to 100% are possible. Here are some tips to minimize the effects of preferred orientation.


The thickness of the powder should be great enough to prevent the beam from passing through to the substrate below. The generally accepted reduction of the beam intensity is about 1/1000th of the initial beam intensity. This condition is termed "infinite thickness".
Examples of powder thickness (µm) required for attenuation of a CuKα beam to 0.01 and 0.001 times incident intensity as a function of 2θ
|
Attenuation |
factor |
|
0.01 |
|
|
|
0.001 |
|
|
||||
|
|
Porosity |
solid |
0.9 |
0.8 |
0.7 |
solid |
0.9 |
0.8 |
0.7 |
||||
|
|
4° |
14.1 |
15.7 |
17.7 |
20.2 |
21.2 |
23.6 |
26.5 |
30.3 |
||||
|
Gibbsite |
20° |
70.4 |
78.2 |
88.0 |
100.5 |
105.6 |
117.3 |
131.9 |
150.8 |
||||
|
µ=56.8 cm-1 |
40° |
138.6 |
154.0 |
173.3 |
198.0 |
207.9 |
231.0 |
259.9 |
297.0 |
||||
|
|
80° |
261.0 |
289.0 |
326.0 |
372.0 |
391.0 |
434.0 |
488.0 |
559.0 |
||||
|
|
4° |
8.3 |
9.3 |
10.4 |
11.9 |
12.5 |
13.9 |
15.6 |
17.8 |
||||
|
Quartz |
20° |
41.4 |
46.0 |
51.8 |
59.2 |
62.2 |
69.1 |
77.7 |
88.8 |
||||
|
µ=96.5 cm-1 |
40° |
81.6 |
90.7 |
102.0 |
116.6 |
122.4 |
136.0 |
153.0 |
174.9 |
||||
|
|
80° |
153.4 |
170.5 |
191.8 |
219.2 |
230.1 |
255.7 |
287.6 |
328.7 |
||||
|
|
4° |
0.7 |
0.8 |
0.9 |
1.0 |
1.1 |
1.2 |
1.3 |
1.5 |
||||
|
Hematite |
20° |
3.6 |
4.0 |
4.34 |
5.1 |
5.3 |
5.9 |
6.7 |
7.6 |
||||
|
µ=1124.4 cm-1 |
40° |
7.0 |
7.8 |
8.8 |
10.0 |
10.5 |
11.7 |
13.1 |
15.0 |
||||
|
|
80° |
13.2 |
14.6 |
16.5 |
18.8 |
19.7 |
21.9 |
24.7 |
28.2 |
||||
| |
|||||||||||||
Factors that influence the transparency of a specimen include